How to remove the well plate fixing bracket for MRX A2000
MRX A2000 Well plate fixing bracket Removal Procedure
#TS-MR1036-001
Introduction
Procedures for how to remove the MRX A2000 Well plate fixing bracket
Preparation
1. MRX A2000
2. (-) Screwdriver
Procedure
1. Open the well plate cover on the front of the equipment in the direction of the arrow as shown in the figure.

2. When you open the well plate cover on the front side, loosen the two screws shown in the picture by turning them counterclockwise using a (-) screwdriver.

3. As shown in the figure, grab the center part of the Well plate fixing bracket and pull it out while lifting it to the front to separate and remove it.

K LAB CO.,LTD.
Phone: +82-42-932-7586󠇌~8
Fax: +82-42-932-7589
E-mail: service@klab.im
Previous
Next
